흉부CT 검사시 Scout과 Helical 관전압 크기 변화에 따른 선량 평가 = The dose assessment due to a change in the helical scan and scout tube voltage on chest CT examination
저자
강태욱(Tae Wook Kang) ; 신상욱(Sang Wook Sin) ; 김규하(Kyu Ha Kim) ; 서정준(Jeong Jun Seo) ; 정종성(Jong Seong Jeong) ; 김명구(Myeong Goo Kim)
발행기관
학술지명
대한CT영상기술학회지(Journal of Korean society of cumputed tomographic technology)
권호사항
발행연도
2014
작성언어
-주제어
KDC
510
등재정보
KCI등재후보
자료형태
학술저널
발행기관 URL
수록면
173-179(7쪽)
제공처
목적 : 흉부CT 검사에서 자동관전류변조기법(ATCM)을 적용한 경우 scout의 관전압과 helical scan 관전압의 크기를 변화시켰을 때 CTDIvol(volume CT dose index)을 통한 선량의 변화를 확인하고자 하였다.
대상 및 방법 : CT장치는 GE社의 Discovery CT 750HD모델이고, ATCM 적용을 위해 tissue-equivalent material로 구성된 Rando Chest Phantom 을 사용하였다.Scout의 노출조건 중 하나인 관전압을 80, 100, 120, 140 kV로 변화시키고 관전류는 10 mA로 고정하여 ATCM을 적용하였다. 실제 흉부CT 검사 시 helical scan 관접압을 80, 100, 120, 140 kV로 변화시켰을 때, 관전류는 고정관전류 200 mA(group A)와 AutomA+SmartmA 10~700 mA(group B), AutomA+SmartmA 100~300 mA(group C)로 구분하여 5회 반복 측정 후 장비에서 모니터링 되는 CTDIvol값을 평가하였다. PASW Statistics v18.0을 이용하여 유의성과 상관분석을 실시하였다.
결과 : Group A에서 CTDIvol은 scout 관전압의 변화에 영향을 받지 않았고 helicalscan 관전압이 80 kV에서 140 kV로 증가함에 따라 4배 이상 증가하였다(r=0.997, p<0.05).
Group B에서 scout이 일정한 경우 helicalscan 관전압이 증가함에 따라 CTDIvol은 최대 19% 감소 하였고(r=-0.475, p<0.05), helicalscan 관전압이 동일한 경우 scout 관전압 증가에 따라 CTDIvol은 최대 17% 감소하였다(r=-0.677, p<0.05).
Group C에서 scout이 일정한 경우 helicalscan 관전압이 증가함에 따라 CTDIvol은 최대 74% 증가하였고(r=0.878, p<0.05), helicalscan 관전압이 동일한 경우 scout 관전압 증가에 따라 CTDIvol은 최대 8% 감소하였다(r=-0.223, p<0.05).
결론 : ATCM을 적용하지 않는 경우 scout 관전압의 변화는CTDIvol에 영향을 주지 않았고, helicalscan 관전압과 CTDIvol은 비례관계를 보였다. ATCM을 적용하여 mA range(10~700mA)를 제한하지 않은 경우 scout과 helicalscan 관전압이 증가할수록 CTDIvol은 감소하였고, mA range(100~300 mA)를 제한한 경우는 scout 관전압은 증가하고 helicalscan 관전압이 감소할수록 CTDIvol은 감소하였다.
Purpose : If you apply the automatic tube current modulation technique (ATCM) in chest CT examination,it has tried to request the change of dose through the CTDIvol when changing the size of the helical scan tube voltage and scout tube voltage.
Materials and methods : CT device is Discovery CT 750HD model GE Co., Using the Rando Chest Phantom consistsequivalent material - tissue for applying ATCM.
Was varied in 80, 100, 120, 140 kV tube voltage, one of the exposure conditions of scout, the tube current was applied ATCM to be fixed to 10 mA.Chest CT scan at the time the actual contact pressure tube for 80, 100, 120, 140 kV was changed by, the tube current is fixed tube current 200 mA(group A) and AutomA + SmartmA 10 ~ 700 mA(group B), AutomA + SmartmA 100 ~ 300 mA(group C) are repeated five times by dividing the measured value was evaluated CTDIvol.
Results : CTDIvol is not affected by changes in the scout tube voltage from Group A, helical scan study tube voltage increases to 140 kV at 80kV, was increased by a factor of four or more(r = 0.997, p <0.05).
In certain cases, was reduced up to 19% is CTDIvolhelicalscan tube voltage increases is scout in Group B(r=-0.475, p<0.05), same case, helicalscan tube voltage is decreased up to 17% is CTDIvol with the increase of tube voltage scout(r=-0.677, p<0.05).
In certain cases, an increase of 74% up to CTDIvolhelical scan tube voltage increases is scout in Group C(r=0.878, p<0.05), same case, helicalscan tube voltage is decreased up to 8% is CTDIvol with the increase of tube voltage scout(r=-0.223, p<0.05).
Conclusion : If not applied ATCM, change of scout tube voltage does not affect the CTDIvol, CTDIvol a tube voltage helicalstudy showed a proportional relationship. If you do not have a limit (10 ~ 700mA) mA range by applying the ATCM, CTDIvol decreased as tube voltage helicalscan and scout is increased. If you have limited (100 ~ 300 mA) mA range, CTDIvol was decreased as the scout tube voltage is increased, helicalscan tube voltage is reduced.
분석정보
서지정보 내보내기(Export)
닫기소장기관 정보
닫기권호소장정보
닫기오류접수
닫기오류 접수 확인
닫기음성서비스 신청
닫기음성서비스 신청 확인
닫기이용약관
닫기학술연구정보서비스 이용약관 (2017년 1월 1일 ~ 현재 적용)
학술연구정보서비스(이하 RISS)는 정보주체의 자유와 권리 보호를 위해 「개인정보 보호법」 및 관계 법령이 정한 바를 준수하여, 적법하게 개인정보를 처리하고 안전하게 관리하고 있습니다. 이에 「개인정보 보호법」 제30조에 따라 정보주체에게 개인정보 처리에 관한 절차 및 기준을 안내하고, 이와 관련한 고충을 신속하고 원활하게 처리할 수 있도록 하기 위하여 다음과 같이 개인정보 처리방침을 수립·공개합니다.
주요 개인정보 처리 표시(라벨링)
목 차
3년
또는 회원탈퇴시까지5년
(「전자상거래 등에서의 소비자보호에 관한3년
(「전자상거래 등에서의 소비자보호에 관한2년
이상(개인정보보호위원회 : 개인정보의 안전성 확보조치 기준)개인정보파일의 명칭 | 운영근거 / 처리목적 | 개인정보파일에 기록되는 개인정보의 항목 | 보유기간 | |
---|---|---|---|---|
학술연구정보서비스 이용자 가입정보 파일 | 한국교육학술정보원법 | 필수 | ID, 비밀번호, 성명, 생년월일, 신분(직업구분), 이메일, 소속분야, 웹진메일 수신동의 여부 | 3년 또는 탈퇴시 |
선택 | 소속기관명, 소속도서관명, 학과/부서명, 학번/직원번호, 휴대전화, 주소 |
구분 | 담당자 | 연락처 |
---|---|---|
KERIS 개인정보 보호책임자 | 정보보호본부 김태우 | - 이메일 : lsy@keris.or.kr - 전화번호 : 053-714-0439 - 팩스번호 : 053-714-0195 |
KERIS 개인정보 보호담당자 | 개인정보보호부 이상엽 | |
RISS 개인정보 보호책임자 | 대학학술본부 장금연 | - 이메일 : giltizen@keris.or.kr - 전화번호 : 053-714-0149 - 팩스번호 : 053-714-0194 |
RISS 개인정보 보호담당자 | 학술진흥부 길원진 |
자동로그아웃 안내
닫기인증오류 안내
닫기귀하께서는 휴면계정 전환 후 1년동안 회원정보 수집 및 이용에 대한
재동의를 하지 않으신 관계로 개인정보가 삭제되었습니다.
(참조 : RISS 이용약관 및 개인정보처리방침)
신규회원으로 가입하여 이용 부탁 드리며, 추가 문의는 고객센터로 연락 바랍니다.
- 기존 아이디 재사용 불가
휴면계정 안내
RISS는 [표준개인정보 보호지침]에 따라 2년을 주기로 개인정보 수집·이용에 관하여 (재)동의를 받고 있으며, (재)동의를 하지 않을 경우, 휴면계정으로 전환됩니다.
(※ 휴면계정은 원문이용 및 복사/대출 서비스를 이용할 수 없습니다.)
휴면계정으로 전환된 후 1년간 회원정보 수집·이용에 대한 재동의를 하지 않을 경우, RISS에서 자동탈퇴 및 개인정보가 삭제처리 됩니다.
고객센터 1599-3122
ARS번호+1번(회원가입 및 정보수정)