액체크로마토그래피-탠덤질량분석법 기반 혈장 메타네프린과 노르메타네프린 정량검사를 위한 전처리자동화법 개발 및 평가 = Development and Validation of an Automated Sample Pretreatment Method for Liquid Chromatography-Tandem Mass Spectrometry Based Quantitation of Plasma Metanephrine and Normetanephrine
저자
Cho Sung-Eun (Department of Endocrine Substance Analysis Center (ESAC), Yongin, KoreaGC Labs, Yongin; GC Labs, Yongin, Korea) ; Han Jungsun (Department of Endocrine Substance Analysis Center (ESAC), Yongin, KoreaGC Labs, Yongin; GC Labs, Yongin, Korea) ; Kim Eunsung (Department of Endocrine Substance Analysis Center (ESAC), Yongin, KoreaGC Labs, Yongin; GC Labs, Yongin, Korea) ; Lee Jun Hyung (Department of Endocrine Substance Analysis Center (ESAC), Yongin, KoreaGC Labs, Yongin; GC Labs, Yongin, Korea) ; Yi Ahram (Department of Endocrine Substance Analysis Center (ESAC), Yongin, KoreaGC Labs, Yongin; GC Labs, Yongin, Korea) ; Lee Sang Gon (GC Labs, Yongin; GC Labs, Yongin, Korea) 연구자관계분석
발행기관
학술지명
권호사항
발행연도
2025
작성언어
Korean
주제어
등재정보
KCI등재
자료형태
학술저널
수록면
214-222(9쪽)
DOI식별코드
제공처
배경: 최근 정확도와 정밀도가 높은 LC-MS/MS 검사 수행건수가 증가함에 따라 검체 전처리 방법도 점점 더 복잡하고 어려워지고 있다. 따라서 수기법으로 이러한 전처리를 수행해야 하는 분석자 의 부담도 증가하고 있다. 이에, 우리는 혈장 메타네프린과 노르메 타네프린의 LC-MS/MS 기반 정량검사의 전처리 단계 자동화를 위 하여 Janus G3 Automated Workstations (Revvity) 기기를 도입하 고, 전처리자동화법을 구현하고 개발했다.
방법: 검사법 세팅 완료 후 정밀도(precision), 정확도(accuracy), 직 선성(liniearity), 하한정량한계(lower limit of quantitation, LLOQ), 잔효(carryover) 등의 검사법 성능을 평가하였고, 전처리수기법과 전처리자동화법의 검사법간 비교도 수행하였다.
결과: 반복정밀도와 검사실내정밀도의 변이계수는 메타네프린과 노르메타네프린 분석시 10% 미만으로 모두 허용 가능했고 , 정확도 편향 결과도 5% 미만으로 모두 허용 가능했다. 검증된 직선성 범위 는 메타네프린 0.02–95.07 nmol/L, 노르메타네프린 0.02–102.34 nmol/L였다. 전처리자동화법의 전처리 시간은 45분으로 초보자에 의해 수행된 전처리수기법 소요시간 172분과 숙련자에 의해 수행된 전처리수기법 소요시간 76분에 비하여 유의하게 짧았다. 전처리 수기법과 전처리자동화법을 이용한 LC-MS/MS 측정값 간의 비교 결과, 메타네프린의 경우, 상관계수 0.9958, 기울기 1.000 (1.000− 1.031), 절편 0.000 (-0.006–0.000)였고, 노르메타네프린의 경우, 상관 계수 0.9976, 기울기 1.000 (0.972−1.005), 절편 0.000 (-0.002–0.022) 로 우수하였다.
결론: Janus G3 Automated Workstations를 이용한 전처리자동화 법은 메타네프린과 노르메타네프린을 정량 분석하는 LC-MS/MS 의 전처리법에 유용한 성능을 보였다.
Background: As the number of LC-MS/MS tests with high accuracy and precision have increased, sample pretreatment methods have also be- come more complex, thus making them more difficult for analysts to perform them manually. To overcome this, we implemented Janus G3 Auto- mated Workstations (Revvity, Turku, Finland) and developed an automated sample pretreatment method for LC-MS/MS based quantitation of plas- ma metanephrine and normetanephrine in this study.
Methods: The feasibility and analytical performances of the method including precision, accuracy, linearity, and carryover, were analyzed and compared with that of the manual sample preparation method.
Results: The coefficients of variations for repeatability and within-laboratory precision were less than 10% for metanephrine and normetaneph- rine assays, meeting acceptable criteria. Accuracy bias values were less than 5%, also within acceptable limits. The verified linear ranges were 0.02 to 95.07 nmol/L for metanephrine and 0.02 to 102.34 nmol/L for normetanephrine. The pretreatment step alone using the Janus G3 Automat- ed Workstations took 45 minutes, significantly shorter than the manual pretreatment method. The comparison between manual and automated pretreatment methods of LC-MS/MS measurement showed excellent agreement, with the following results: a correlation coefficient of 0.9958, slope of 1.000 (1.000–1.031), and intercept of 0.000 (-0.006–0.000) for plasma metanephrine, and 0.9976, 1.000 (0.972–1.005), and 0.000 (-0.002– 0.022) for plasma normetanephrine, respectively.
Conclusions: This automated sample pretreatment method with Janus G3 Automated Workstations showed acceptable performances for pre- treatment steps of LC-MS/MS quantitation regarding both metanephrine and normetanephrine. This method offers a beneficial solution for ana- lysts by reducing labor.
분석정보
서지정보 내보내기(Export)
닫기소장기관 정보
닫기권호소장정보
닫기오류접수
닫기오류 접수 확인
닫기음성서비스 신청
닫기음성서비스 신청 확인
닫기이용약관
닫기학술연구정보서비스 이용약관 (2017년 1월 1일 ~ 현재 적용)
| 주요 개정내역 | 변경 사유 |
|---|---|
| · 수탁업체 콘소시엄 기관명 및 위탁기간 명시 | · 제6조(개인정보 처리업무의 위탁) 구체화 |
한국교육학술정보원은 정보주체의 자유와 권리 보호를 위해 「개인정보 보호법」 및 관계 법령이 정한 바를 준수하여, 적법하게 개인정보를 처리하고 안전하게 관리하고 있습니다. 이에 「개인정보 보호법」 제30조에 따라 정보주체에게 개인정보 처리에 관한 절차 및 기준을 안내하고, 이와 관련한 고충을 신속하고 원활하게 처리할 수 있도록 하기 위하여 다음과 같이 개인정보 처리방침을 수립·공개합니다.
주요 개인정보 처리 표시(라벨링)
목 차
제1조(개인정보의 처리 목적)
제2조(개인정보의 처리 및 보유 기간)
제3조(처리하는 개인정보의 항목)
제4조(개인정보파일 등록 현황)
제5조(개인정보의 제3자 제공)
제6조(개인정보 처리업무의 위탁)
제7조(개인정보의 파기 절차 및 방법)
제8조(정보주체와 법정대리인의 권리·의무 및 그 행사 방법)
제9조(개인정보의 안전성 확보조치)
제10조(개인정보 자동 수집 장치의 설치·운영 및 거부)
제11조(개인정보 보호책임자)
제12조(개인정보의 열람청구를 접수·처리하는 부서)
제13조(정보주체의 권익침해에 대한 구제방법)
제14조(추가적 이용·제공 판단기준)
제15조(개인정보 처리방침의 변경)
제1조(개인정보의 처리 목적)
제2조(개인정보의 처리 및 보유 기간)
3년
또는 회원탈퇴시까지5년
(「전자상거래 등에서의 소비자보호에 관한3년
(「전자상거래 등에서의 소비자보호에 관한2년
이상(개인정보보호위원회 : 개인정보의 안전성 확보조치 기준)
제3조(처리하는 개인정보의 항목)
제4조(개인정보파일 등록 현황)
개인정보파일 검색(privacy.go.kr)| 개인정보파일의 명칭 | 운영근거 / 처리목적 | 개인정보파일에 기록되는 개인정보의 항목 |
보유기간 | |
|---|---|---|---|---|
| 학술연구정보서비스 이용자 가입정보 | 한국교육학술정보원법 정보추제 동의 | 필수 | ID, 비밀번호, 성명, 생년월일, 신분(직업구분), 이메일, 소속분야, 웹진메일 수신동의 여부 | 3년 또는 탈퇴시 |
| 선택 | 소속기관명, 소속도서관명, 학과/부서명, 학번/직원번호, 휴대전화, 주소 | |||
제5조(개인정보의 제3자 제공)
제6조(개인정보 처리업무의 위탁)
제7조(개인정보의 파기 절차 및 방법)
제8조(정보주체와 법정대리인의 권리·의무 및 그 행사 방법)
제9조(개인정보의 안전성 확보조치)
제10조(개인정보 자동 수집 장치의 설치·운영 및 거부)
제11조(개인정보 보호책임자)
| 구분 | 담당자 | 연락처 |
|---|---|---|
| KERIS 개인정보 보호책임자 | 정보보호본부 안재호 |
- 이메일 : jinuk@keris.or.kr - 전화번호 : 053-714-0158 - 팩스번호 : 053-714-0195 |
| KERIS 개인정보 보호담당자 | 개인정보보호부 송진욱 | |
| RISS 개인정보 보호책임자 | 교육학술데이터본부 정광훈 |
- 이메일 : giltizen@keris.or.kr - 전화번호 : 053-714-0149 - 팩스번호 : 053-714-0194 |
| RISS 개인정보 보호담당자 | 학술진흥부 길원진 |
제12조(개인정보의 열람청구를 접수·처리하는 부서)
제13조(정보주체의 권익침해에 대한 구제방법)
제14조(추가적인 이용ㆍ제공 판단기준)
제15조(개인정보 처리방침의 변경)
자동로그아웃 안내
닫기인증오류 안내
닫기귀하께서는 휴면계정 전환 후 1년동안 회원정보 수집 및 이용에 대한
재동의를 하지 않으신 관계로 개인정보가 삭제되었습니다.
(참조 : RISS 이용약관 및 개인정보처리방침)
신규회원으로 가입하여 이용 부탁 드리며, 추가 문의는 고객센터로 연락 바랍니다.
- 기존 아이디 재사용 불가
휴면계정 안내
RISS는 [표준개인정보 보호지침]에 따라 2년을 주기로 개인정보 수집·이용에 관하여 (재)동의를 받고 있으며, (재)동의를 하지 않을 경우, 휴면계정으로 전환됩니다.
(※ 휴면계정은 원문이용 및 복사/대출 서비스를 이용할 수 없습니다.)
휴면계정으로 전환된 후 1년간 회원정보 수집·이용에 대한 재동의를 하지 않을 경우, RISS에서 자동탈퇴 및 개인정보가 삭제처리 됩니다.
고객센터 1599-3122
ARS번호+1번(회원가입 및 정보수정)