Development of 3-D Stereotactic Localization System and Rediation Measurement for Stereotactic Radiosurgery = 방사선수술을 위한 3차원 정위 시스템 및 방사선량 측정 시스템 개발
저자
Suh,Tae-Suk (Catholic University) ; Suh,Doug-Young (Kyung Hee University) ; Park,Sung-Hun (Kun Kuk University) ; Jang,Hong-Seok (Catholic University) ; Choe,Bo-Young (Catholic University) ; Yoon,Sei-Chul (Catholic University) ; Shinn,Kyung-Sub (Catholic University) ; Bahk,Yong-Whee (Catholic University) ; Kim,Il-Hwan (Seoul National University) ; Kang,Wee-Sang (Seoul National University) ; Ha,Sung-Whan (Seoul National University) ; Park,Charn-Il (Seoul National University)
발행기관
학술지명
권호사항
발행연도
1995
작성언어
Korean
주제어
KDC
559.17
등재정보
KCI등재
자료형태
학술저널
발행기관 URL
수록면
25-36(12쪽)
제공처
소장기관
본 연구는 방사선 수술시 필수적으로 이용되는 환자의 정위적 데이터 및 방사선 측정 데이터를 효율적으로 처리하여 종래의 시행착오적 방식을 통한 수술 계획상의 오류 및 부정확성을 지양하고 능률적이고 효율적인 수술이 될 수 있도록 하는데 목적이 있다.
정위시스템 개발을 위하여 Fisher 프레임이 사용되었으며, 뇌혈관 촬영 및 CT의 정위적 방법에 의하여 병변의 위치를 3 차원적으로 정확히 얻을 수 있는 알고리듬을 개발하였다. 정위시스템 과정은 컴퓨터에 의해 쉽게 운열될 수 있도록 프로그램화하였으며, 환자 영상을 컴퓨터에 입력하는 과정, 입력된 영상으로부터 병소 및 위치측정 보조기구 (localizer)에 대한 정보를 추출하는 과정, 입력된 영상의 위치 정보로부터 병소의 3 차원적 정위좌표로 변환하는 과정 등으로 나누었다. 좌표 변환 계산은 실시산에 구해졌으며, CT 및 혈관촬영술의 정위적 방법에 의하여 얻어진 병소의 3차원 좌표를 서로 비교한 결과 오차는 통상적으로 인정되는 2 mm 이내였다.
NEC 6 MVX 선형 가속기를 사용하여 소조사면 선량을 측정하였다. 사용된 측정기로는 전리함(ion chamber), 필름, 다이오드 등이며, 측정된 값은 출력 계수, 깊이선량률 (PDD), 최대조직선량비(TMR), 축선량비 (OAR) 등이다. 측정기의 종류에 따라, 측정치가 조금씩 상이했으나, 전반적으로 같은 경향을 보여 줌으로 써 측정치의 신뢰도를 확인할 수 있었다. 제작된 구형 물팬톰과 필름을 이용하여 표준 방사선 치료시 예측했던 선량 분포가 실제 나오는가를 실험을 통하여 확인하였으며, 예측한 선량 값을 얻을 수 있었다.
본 연구를 통하여 개인용컴퓨터에 의해 병소의 정위좌표를 정확히 얻을 수 있는 프로그램이 개발되었고, 소조사면을 간단하고 편리하게 잴 수 있는 방법을 개발함으로써 방사선 수술에 많은 도움이 될 것으로 사료된다.
The purpose of this research is to develop stereotactic localization and radiation measurement system for the efficient and precise radiosurgery.
The algorithm to obtain a 3-D stereotactic coordinates of the target has been developed using a Fisher CT or angio localization. The procedure of stereotactic localization was programmed with PC computer, and consists of three steps: (1) transferring patient images into PC; (2) marking the position of target and reference points of the localizaer from the patient image; (3) computing the stereotactic 3-D coordinates of target associated with position information of localizer. Coordinate transformation was quickly done on a real time base. The difference of coordinates computed from between Angio and CT localization method was within 2 mm, which could be generally accepted for the reliability of the localization system developed.
We measured dose distribution in small fields of NEC 6 MVX linear accelerator using various detector; ion chamber, film, diode. Specific quantities measured include output factor, percent depth dose (PDD), tissue maximum ratio (TMR), off-axis ratio (OAR). There was small variation of measured data according to the different kinds of detectors used. The overall trends of measured beam data were similar enough to rely on our measurement. The measurement was performed with the use of hand-made spherical water phantom and film for standard arc set-up. We obtained the dose distribution as we expected.
In conclusion, PC-based 3-D stereotactic localization system was developed to determine the stereotactic coordinate of the target. A convenient technique for the small field measurement was demonstrated. Those methods will be much helpful for the stereotactic radiosurgery.
분석정보
서지정보 내보내기(Export)
닫기소장기관 정보
닫기권호소장정보
닫기오류접수
닫기오류 접수 확인
닫기음성서비스 신청
닫기음성서비스 신청 확인
닫기이용약관
닫기학술연구정보서비스 이용약관 (2017년 1월 1일 ~ 현재 적용)
학술연구정보서비스(이하 RISS)는 정보주체의 자유와 권리 보호를 위해 「개인정보 보호법」 및 관계 법령이 정한 바를 준수하여, 적법하게 개인정보를 처리하고 안전하게 관리하고 있습니다. 이에 「개인정보 보호법」 제30조에 따라 정보주체에게 개인정보 처리에 관한 절차 및 기준을 안내하고, 이와 관련한 고충을 신속하고 원활하게 처리할 수 있도록 하기 위하여 다음과 같이 개인정보 처리방침을 수립·공개합니다.
주요 개인정보 처리 표시(라벨링)
목 차
3년
또는 회원탈퇴시까지5년
(「전자상거래 등에서의 소비자보호에 관한3년
(「전자상거래 등에서의 소비자보호에 관한2년
이상(개인정보보호위원회 : 개인정보의 안전성 확보조치 기준)개인정보파일의 명칭 | 운영근거 / 처리목적 | 개인정보파일에 기록되는 개인정보의 항목 | 보유기간 | |
---|---|---|---|---|
학술연구정보서비스 이용자 가입정보 파일 | 한국교육학술정보원법 | 필수 | ID, 비밀번호, 성명, 생년월일, 신분(직업구분), 이메일, 소속분야, 웹진메일 수신동의 여부 | 3년 또는 탈퇴시 |
선택 | 소속기관명, 소속도서관명, 학과/부서명, 학번/직원번호, 휴대전화, 주소 |
구분 | 담당자 | 연락처 |
---|---|---|
KERIS 개인정보 보호책임자 | 정보보호본부 김태우 | - 이메일 : lsy@keris.or.kr - 전화번호 : 053-714-0439 - 팩스번호 : 053-714-0195 |
KERIS 개인정보 보호담당자 | 개인정보보호부 이상엽 | |
RISS 개인정보 보호책임자 | 대학학술본부 장금연 | - 이메일 : giltizen@keris.or.kr - 전화번호 : 053-714-0149 - 팩스번호 : 053-714-0194 |
RISS 개인정보 보호담당자 | 학술진흥부 길원진 |
자동로그아웃 안내
닫기인증오류 안내
닫기귀하께서는 휴면계정 전환 후 1년동안 회원정보 수집 및 이용에 대한
재동의를 하지 않으신 관계로 개인정보가 삭제되었습니다.
(참조 : RISS 이용약관 및 개인정보처리방침)
신규회원으로 가입하여 이용 부탁 드리며, 추가 문의는 고객센터로 연락 바랍니다.
- 기존 아이디 재사용 불가
휴면계정 안내
RISS는 [표준개인정보 보호지침]에 따라 2년을 주기로 개인정보 수집·이용에 관하여 (재)동의를 받고 있으며, (재)동의를 하지 않을 경우, 휴면계정으로 전환됩니다.
(※ 휴면계정은 원문이용 및 복사/대출 서비스를 이용할 수 없습니다.)
휴면계정으로 전환된 후 1년간 회원정보 수집·이용에 대한 재동의를 하지 않을 경우, RISS에서 자동탈퇴 및 개인정보가 삭제처리 됩니다.
고객센터 1599-3122
ARS번호+1번(회원가입 및 정보수정)