CT검사에서의 환자선량 표준화에 대한 연구 : Standard Dose Abdominal CT = A Study on the standardization of radiation dose in CT : Stand Dose Abdominal CT
저자
강승원(Seung Won Kang) ; 정용환(Yong Hwan Jung) ; 남태현(Tae Hyen Nam) ; 김태은(Tae Eun Kim) ; 권순안(Sun Ahn Gwon) ; 민관홍(Kwan Hong Min)
발행기관
학술지명
대한CT영상기술학회지(Journal of Korean society of cumputed tomographic technology)
권호사항
발행연도
2016
작성언어
-주제어
KDC
512.151
등재정보
KCI등재후보
자료형태
학술저널
발행기관 URL
수록면
49-57(9쪽)
제공처
목적 : 1phase Standard Dose abdominal CT(이하 SDA CT) 검사에서 환자가 받는 선량의 차이가 최소화되도록, 각 장비 별, 검사자 별로 선량 calibration을 진행하여 선량의 차이를 최소화하고 표준화하고자 한다. 대상 및 방법 : 2015년 1월부터 2015년 11월까지 본원에 내원한 환자중 SDA CT검사를 시행한 성인(만16세 이상)환자 3000명을 대상으로 하였다. CT Scanner로 사용한 장비는 64-slice MDCT(Brilliance CT, Philips Medical Systems, The Netherlands) 3대, 256-slice MDCT(2561CT, Philips Medical Systems, The Netherlands) 2대, 총 5대의 MDCT(이하, A, B, C. D, E)를 사용하였다. 실험 data의 수집은 선량관리프로그램(이하, Dose track)을 이용하였고 분석은 MedCalc(Version 10.0)을 이용하였으며 Box & whisker plot chart를 이용하여 median Dose Length Product (이하, median DLP)을 비교분석하였다. 본 논문에서는 SDA CT의 median DLP의 목표치를 530mGy*cm로 정하였고, 오차 허용범위를 ±10% 로 하였다. (이하 목표기준 DLP값)실험방법은 크게 세 가지로 구분하여 진행하였다. 첫 번째, 장비 선량 Calibration에서는 각 장비 별로 현재 사용하고 있는 Protocol의SDA CT의 median DLP값을 알아보고, 목표 기준 DLP값과 차이가 있을 시 오차 허용범위 안에 들어오도록 공식을 적용하여 목표 기준 mAs값을 변경하였다. 두 번째 검사자 별 선량 Calibration 에서는 10명의 검사자 별로 SDA CT의 median DLP값을 알아보고, 목표기준 DLP값과 차이가 있을 시 오차 허용범위 안에 들어오도록 protocol 수정하여 교육하였다. 세 번째, 영상평가는 영의학과 판독의 2명을 대상으로 후향적 평가(Retrospective Study)로 4점척도로 진행하였다. 결과 : 장비 선량 Calibration에서는 5대의 장비 A, B, C, D, E의 median DLP값은 490.2mGy*cm ~ 626.6mGy*cm로 나타났다. D와 E 2대의 장비에서 median DLP값은 각각 10.5%. 18.2% 차이로 목표 허용 범위에 충족되지 않았다. 공식을 이용하여 D와 E의 현재 기준 mAs값을 340mAs와 320mAs로 변경한 후, 각 장비의 median DLP값을 확인한 결과 4B6.7mGy*cm ~ 568.1mGy*cm로 나타났으며, D와 E를 포함한 모든 장비가 허용 범위 내에 충족되었다. 검사자 선량 Calibration에서는 10명의 검사자 RT1~RT10의 median DLP값은 491.SmGy*cm ~ 601.75mGy*cm로 나타났다. 검사자 RT1과 RT10,2명의 검사자 median DLP값은 각각 12.5%, 13.5% 차이로 목표 허용범위에 충족되지 않았다. Protocol 수정 후 median DLP값을 확인한 결과 498.5mGy*cm ~ 550.3mGy*cm로 나타났으며, RT1, RT10을 포함한 모든 검사자가 허용 범위 내에 충족되었다. 판독의 2명을 대상으로 진행한 영상평가에서는, 각각 3.5, 3.7 점으로 모든 영상이 판독 가능했다. 결론 : 제조사별, 모델 별, 또는 detector나 tube의 노후화에 따라 선량차이가 날 수 있으며, 자동 노출 제어 장치의 사용과 같은 다양한 조건으로 인해 선량의 차이가 나타날 수 밖에 없지만, 본 논문의 선량 calibration 방법처럼 적용하기 위해 노력한다면 동일한 검사에서 선량의 차이를 최소화 할 수 있을 것으로 사료된다.
더보기Purpose : The purpose of this study is to minimize the dose difference by conducting a calibration of each device by the inspectors and standardize for the radiation dose that the patient receives in 1 phase Standard Dose abdominal CT(SDA CT) Materials and Methods: It targeted at the 3,000 adult patients (over 16 years) who underwent SDA CT examination until November 2015 from January 2015. 64-slice MDCT (Brilliance CT, Philips Medical Systems. The Netherlands) 3 units, 256-slice MDCT (256iCT, Philips Medical Systems, The Netherlands) 2 units, a total five MDCT was used. Experimental data was collected by using Dose track and analyzed by using Medcalc(Version 10.0) and median Dose Length Product (median DLP) was compared and analyzed by using Box & whisker plot chart. In this paper, the target value of the median DLP of SDA CT was 530mGy * cm, the error tolerance range was± 10%. (Target standard DLP value) Experimental procedure was preceded by three major. First. we researched median DLP value of the current protocolin the case of Calibration of the dose for each device and changed a targeted mAs value by applying the formula to get in an error tolerance range when there was a difference between DLP values which was targeted at. Secondly, we researched the median DLP of SDA CT for each 10 inspectors and when finding the differences, inspectors were trained modified protocol to get in tolerance error range. Third, the image evaluation was preceded at 4-point scale of the retrospective evaluation (Retrospective Study) by 2 radiologists. Result : In equipment Dose Calibration, five devices A, B, C, D, E’s median DLP value were 490.2mGy * cm~626.6mGy *cm and median DLP value of D, E were 10.5%, 18.2% respectively. These were not met on the target range. Using the formula, after changing current mAs of D and E to 340mAs, 320mAs, the results showed 486.7mGy * cm~568.1 mGy * and it met on the reference range. In 10 RT inspector dose Calibration, the median DLP value appeared 491.5mGy * cm~601.75mGy * cm. Median DLP value of Inspector RTl and RT10 showed 12.5%, 13.5% respectively. These were not met on the target range. Except RT1, RT10, Median DLP values were met on the reference target range after changing the Protocol to 498.5mGy*cm ~ 550.3mGy*cm. In the image evaluation by 2 radiologists, all of the images was decipherable. Conclusion : There was a dose difference between the equipment brands, outdated model and detector were able to make a dose difference. Even variety of conditions of using automatic exposure control device can make a dose difference, but if there is the effort to apply the method of this paper, it is considered to be able to minimize the dose differences between the same examinations.
더보기서지정보 내보내기(Export)
닫기소장기관 정보
닫기권호소장정보
닫기오류접수
닫기오류 접수 확인
닫기음성서비스 신청
닫기음성서비스 신청 확인
닫기이용약관
닫기학술연구정보서비스 이용약관 (2017년 1월 1일 ~ 현재 적용)
학술연구정보서비스(이하 RISS)는 정보주체의 자유와 권리 보호를 위해 「개인정보 보호법」 및 관계 법령이 정한 바를 준수하여, 적법하게 개인정보를 처리하고 안전하게 관리하고 있습니다. 이에 「개인정보 보호법」 제30조에 따라 정보주체에게 개인정보 처리에 관한 절차 및 기준을 안내하고, 이와 관련한 고충을 신속하고 원활하게 처리할 수 있도록 하기 위하여 다음과 같이 개인정보 처리방침을 수립·공개합니다.
주요 개인정보 처리 표시(라벨링)
목 차
3년
또는 회원탈퇴시까지5년
(「전자상거래 등에서의 소비자보호에 관한3년
(「전자상거래 등에서의 소비자보호에 관한2년
이상(개인정보보호위원회 : 개인정보의 안전성 확보조치 기준)개인정보파일의 명칭 | 운영근거 / 처리목적 | 개인정보파일에 기록되는 개인정보의 항목 | 보유기간 | |
---|---|---|---|---|
학술연구정보서비스 이용자 가입정보 파일 | 한국교육학술정보원법 | 필수 | ID, 비밀번호, 성명, 생년월일, 신분(직업구분), 이메일, 소속분야, 웹진메일 수신동의 여부 | 3년 또는 탈퇴시 |
선택 | 소속기관명, 소속도서관명, 학과/부서명, 학번/직원번호, 휴대전화, 주소 |
구분 | 담당자 | 연락처 |
---|---|---|
KERIS 개인정보 보호책임자 | 정보보호본부 김태우 | - 이메일 : lsy@keris.or.kr - 전화번호 : 053-714-0439 - 팩스번호 : 053-714-0195 |
KERIS 개인정보 보호담당자 | 개인정보보호부 이상엽 | |
RISS 개인정보 보호책임자 | 대학학술본부 장금연 | - 이메일 : giltizen@keris.or.kr - 전화번호 : 053-714-0149 - 팩스번호 : 053-714-0194 |
RISS 개인정보 보호담당자 | 학술진흥부 길원진 |
자동로그아웃 안내
닫기인증오류 안내
닫기귀하께서는 휴면계정 전환 후 1년동안 회원정보 수집 및 이용에 대한
재동의를 하지 않으신 관계로 개인정보가 삭제되었습니다.
(참조 : RISS 이용약관 및 개인정보처리방침)
신규회원으로 가입하여 이용 부탁 드리며, 추가 문의는 고객센터로 연락 바랍니다.
- 기존 아이디 재사용 불가
휴면계정 안내
RISS는 [표준개인정보 보호지침]에 따라 2년을 주기로 개인정보 수집·이용에 관하여 (재)동의를 받고 있으며, (재)동의를 하지 않을 경우, 휴면계정으로 전환됩니다.
(※ 휴면계정은 원문이용 및 복사/대출 서비스를 이용할 수 없습니다.)
휴면계정으로 전환된 후 1년간 회원정보 수집·이용에 대한 재동의를 하지 않을 경우, RISS에서 자동탈퇴 및 개인정보가 삭제처리 됩니다.
고객센터 1599-3122
ARS번호+1번(회원가입 및 정보수정)