PET/CT에서 재구성 프로그램의 성능 평가 = Evaluation of Image Quality Based on Time of Flight in PET/CT
저자
임정진 ; 윤석환 ; 김종필 ; 남궁식 ; 신성화 ; 윤상혁 ; 김영석 ; 이형진 ; 이홍재 ; 김진의 ; 우재룡 ; Lim, Jung Jin ; Yoon, Seok Hwan ; Kim, Jong Pil ; Nam Koong, Sik ; Shin, Seong Hwa ; Yoon, Sang Hyeok ; Kim, Yeong Seok ; Lee, Hyeong Jin ; Lee, Hong Jae ; Kim, Jin Eui ; Woo, Jae Ryong
발행기관
학술지명
권호사항
발행연도
2012
작성언어
Korean
주제어
자료형태
학술저널
수록면
110-114(5쪽)
제공처
암의 조기검진 및 수술 전후 추적검사에 유용하게 이용되고 있는 PET/CT는 영상의 질을 향상시키기 위하여 기계적인 성능 향상과 더불어 영상 재구성방법도 발전되어 왔다. 본 연구는 Time of Flight (TOF)를 기반으로 한 재구성 프로그램들에 대하여 영상의 질을 평가하고자 한다. Gemini TF, Biograph mCT, Discovery 690을 이용하여 phantom 영상을 동일한 조건으로 2분 동안 영상을 획득 후 Astonish TF, ultraHD PET, SharpIR을 적용한 것과 적용하지 않은 것에 대하여 영상을 재구성하였다. Flangeless Esser PET phantom 의 내부에는 $^{18}F$-FDG 1.11 kBq/ml (30 ${\mu}Ci/ml$)를 채우고 4개의 열소 원통(8, 12, 16, 25 mm)에는 8.88 kBq/ml (240 ${\mu}Ci/ml$)를 채워서 배후 방사능과 열소 원통 방사능의 비율이 1:8이 되도록 제작하였고 triple line phantom의 내부에는 $^{18}F$-FDG 37 MBq (1 mCi)를 채우고 세 개의 line에는 0.37 MBq/ml (100 uCi)를 주입하여 제작하였다. Flangeless Esser PET phantom을 사용한 재구성 영상에서 contrast ratio와 background variability를 구하였고, triple line phantom을 사용한 재구성 영상에서 resolution을 측정하였다. Phantom lid 크기가 8, 12, 16, 25 mm에서의 contrast ratio는 Astonish TF를 적용하지 않은 영상에서 8.69, 12.28, 19.31, 25.80%, 적용한 영상에서는 6.24, 13.24, 19.55, 27.60%, ultraHD PET를 적용하지 않은 영상에서 4.94, 12.68, 22.09, 30.14%, 적용한 영상에서는 4.76, 13.23, 23.72, 31.65%, SharpIR를 적용하지 않은 영상에서 13.18, 17.44, 28.76, 34.67%, 적용한 영상에서는 13.15, 18.32, 30.33, 35.73%로 나타났다. Background variability는 Astonish TF를 적용하지 않은 영상에서 5.51, 5.42, 7.13, 6.28%, 적용한 영상에서는 7.81, 7.94, 6.40, 6.28%, ultraHD PET를 적용하지 않은 영상에서 6.46, 6.63, 5.33, 5.21%, 적용한 영상에서는 6.08, 6.08, 4.45, 4.58%, SharpIR를 적용하지 않은 영상에서 5.93, 4.82, 4.45, 5.09%, 적용한 영상에서는 4.80, 3.92, 3.63, 4.50%로 나타났다. Phantom line 위치가 upper, center, right에서의 resolution은 Astonish TF를 적용하지 않은 영상에서 10.77, 11.54, 9.34 mm, 적용한 영상에서는 9.54, 8.90, 8.88 mm, ultraHD PET를 적용하지 않은 영상에서 7.84, 6.95, 8.32 mm, 적용한 영상에서는 7.51, 6.66, 8.27 mm, SharpIR를 적용하지 않은 영상에서 9.35, 8.69, 8.99 mm, 적용한 영상에서는 9.88, 9.18, 9.00 mm로 나타났다. TOF를 기반으로 하여 영상의 질을 향상시키기 위한 재구성 프로그램 사용 시 전반적으로 영상의 질적 향상이 이루어짐을 알 수 있었다. 또한 제조사별 재구성 프로그램 비교에 대해서는 어느 정도의 결과 값의 차이를 보였지만 이는 제조사별 장비의 특성과 재구성 알고리즘의 차이로 인한 결과라고 생각된다. 따라서 각 병원에서는 영상의 질을 향상시키기 위해 사용되는 재구성 프로그램을 이용함에 있어서 프로그램에 맞는 적절한 재구성 조건을 찾기 위한 추가적인 연구가 필요할 것으로 사료된다.
더보기Purpose : PET/CT is widely used for early checking up of cancer and following up of pre and post operation. Image reconstruction method is advanced with mechanical function. We want to evaluate image quality of each reconstruction program based on time of flight (TOF). Materials and Methods : After acquiring phantom images during 2 minutes with Gemini TF (Philips, USA), Biograph mCT (Siemens, USA) and Discovery 690 (GE, USA), we reconstructed image applied to Astonish TF (Philips, USA), ultraHD PET (Siemens, USA), Sharp IR (GE, USA) and not applied. inside of Flangeless Esser PET phantom (Data Spectrum corp., USA) was filled with $^{18}F$-FDG 1.11 kBq/ml (30 Ci/ml) and 4 hot inserts (8. 12. 16. 25 mm) were filled with 8.88 kBq/ml (240 ${\mu}Ci/ml$) the ratio of background activity and hot inserts activity was 1 : 8. Inside of triple line phantom (Data Spectrum corp., USA) was filled with $^{18}F$-FDG 37 MBq/ml (1 mCi). Three of lines were filled with 0.37 MBq (100 ${\mu}Ci$). Contrast ratio and background variability were acquired from reconstruction image used Flangeless Esser PET phantom and resolution was acquired from reconstruction image used triple line phantom. Results : The contrast ratio of image which was not applied to Astonish TF was 8.69, 12.28, 19.31, 25.80% in phantom lid of which size was 8, 12, 16, 25 mm and it which was applied to Astonish TF was 6.24, 13.24, 19.55, 27.60%. It which was not applied to ultraHD PET was 4.94, 12.68, 22.09, 30.14%, it which was applied to ultraHD PET was 4.76, 13.23, 23.72, 31.65%. It which was not applied to SharpIR was 13.18, 17.44, 28.76, 34.67%, it which was applied to SharpIR was 13.15, 18.32, 30.33, 35.73%. The background variability of image which was not applied to Astonish TF was 5.51, 5.42, 7.13, 6.28%. it which was applied to Astonish TF was 7.81, 7.94, 6.40 6.28%. It which was not applied to ultraHD PET was 6.46, 6.63, 5.33, 5.21%, it which was applied to ultraHD PET was 6.08, 6.08, 4.45, 4.58%. It which was not applied to SharpIR was 5.93, 4.82, 4.45, 5.09%, it which was applied to SharpIR was 4.80, 3.92, 3.63, 4.50%. The resolution of phantom line of which location was upper, center, right, which was not applied to Astonish TF was 10.77, 11.54, 9.34 mm it which was applied to Astonish TF was 9.54, 8.90, 8.88 mm. It which was not applied to ultraHD PET was 7.84, 6.95, 8.32 mm, it which was applied to ultraHD PET was 7.51, 6.66, 8.27 mm. It which was not applied to SharpIR was 9.35, 8.69, 8.99, it which was applied to SharpIR was 9.88, 9.18, 9.00 mm. Conclusion : Image quality was advanced generally while reconstruction program which is based on time of flight was used. Futhermore difference of result compared each manufacture reconstruction program showed up, however this is caused by specification of instrument of each manufacture and difference of reconstruction algorithm. Therefore we need further examination to find out appropriate reconstruction condition while using reconstruction program used for advance of image quality.
더보기분석정보
서지정보 내보내기(Export)
닫기소장기관 정보
닫기권호소장정보
닫기오류접수
닫기오류 접수 확인
닫기음성서비스 신청
닫기음성서비스 신청 확인
닫기이용약관
닫기학술연구정보서비스 이용약관 (2017년 1월 1일 ~ 현재 적용)
학술연구정보서비스(이하 RISS)는 정보주체의 자유와 권리 보호를 위해 「개인정보 보호법」 및 관계 법령이 정한 바를 준수하여, 적법하게 개인정보를 처리하고 안전하게 관리하고 있습니다. 이에 「개인정보 보호법」 제30조에 따라 정보주체에게 개인정보 처리에 관한 절차 및 기준을 안내하고, 이와 관련한 고충을 신속하고 원활하게 처리할 수 있도록 하기 위하여 다음과 같이 개인정보 처리방침을 수립·공개합니다.
주요 개인정보 처리 표시(라벨링)
목 차
3년
또는 회원탈퇴시까지5년
(「전자상거래 등에서의 소비자보호에 관한3년
(「전자상거래 등에서의 소비자보호에 관한2년
이상(개인정보보호위원회 : 개인정보의 안전성 확보조치 기준)개인정보파일의 명칭 | 운영근거 / 처리목적 | 개인정보파일에 기록되는 개인정보의 항목 | 보유기간 | |
---|---|---|---|---|
학술연구정보서비스 이용자 가입정보 파일 | 한국교육학술정보원법 | 필수 | ID, 비밀번호, 성명, 생년월일, 신분(직업구분), 이메일, 소속분야, 웹진메일 수신동의 여부 | 3년 또는 탈퇴시 |
선택 | 소속기관명, 소속도서관명, 학과/부서명, 학번/직원번호, 휴대전화, 주소 |
구분 | 담당자 | 연락처 |
---|---|---|
KERIS 개인정보 보호책임자 | 정보보호본부 김태우 | - 이메일 : lsy@keris.or.kr - 전화번호 : 053-714-0439 - 팩스번호 : 053-714-0195 |
KERIS 개인정보 보호담당자 | 개인정보보호부 이상엽 | |
RISS 개인정보 보호책임자 | 대학학술본부 장금연 | - 이메일 : giltizen@keris.or.kr - 전화번호 : 053-714-0149 - 팩스번호 : 053-714-0194 |
RISS 개인정보 보호담당자 | 학술진흥부 길원진 |
자동로그아웃 안내
닫기인증오류 안내
닫기귀하께서는 휴면계정 전환 후 1년동안 회원정보 수집 및 이용에 대한
재동의를 하지 않으신 관계로 개인정보가 삭제되었습니다.
(참조 : RISS 이용약관 및 개인정보처리방침)
신규회원으로 가입하여 이용 부탁 드리며, 추가 문의는 고객센터로 연락 바랍니다.
- 기존 아이디 재사용 불가
휴면계정 안내
RISS는 [표준개인정보 보호지침]에 따라 2년을 주기로 개인정보 수집·이용에 관하여 (재)동의를 받고 있으며, (재)동의를 하지 않을 경우, 휴면계정으로 전환됩니다.
(※ 휴면계정은 원문이용 및 복사/대출 서비스를 이용할 수 없습니다.)
휴면계정으로 전환된 후 1년간 회원정보 수집·이용에 대한 재동의를 하지 않을 경우, RISS에서 자동탈퇴 및 개인정보가 삭제처리 됩니다.
고객센터 1599-3122
ARS번호+1번(회원가입 및 정보수정)